Per-mass Single Ion Area on an ICP-ToF ====================================== Determining the :term:`single ion area` (SIA) is essential for accurate thresholding of ICP-ToF data. In SPCal we approximate the SIA using a shape parameter, :math:`\sigma`. This value can be determined from ionic data using the methods described in :ref:`Recovery of compound-Poisson-lognormal parameters`. In the SPCal GUI we can determine SIA values for each mass by loading a low concetration ionic standard into the :ref:`Single Ion Distribution Dialog`, found in the **Limit Options Dock**. #. Download the required data file. The ``tof_single_ion`` directory can be found in the `example_2_data.zip `_ archive. #. Start the :ref:`Single Ion Distribution Dialog` Click the *Single Ion Options...* button in the **Compound** tab of the **Limit Options Dock**. This will start the dialog. #. Open the example data file. Once started, the dialog will prompt you to load a file. Load the ``tof_single_ion/run.info`` file. .. _tutorial single ion: .. figure:: ../images/tutorial_single_ion.png :width: 60% :align: center The single ion dialog. The dialog should now look like :numref:`tutorial single ion`, with a histogram of all signals shown on the left and calcualated shapes on the right. #. Check the calculated shape values for anomalies. The shapes are shown as a scatter plot on the right half of the dialog, as in :numref:`tutorial single ion`. Red points have been excluded due to low or high zero counts (preventing calculation of :math:`\lambda`) or being to far from the mean value. Both silver isotopes (107 and 109) have very high shape values of around 1.2, and are thus excluded. *Left-click* on either point to display the signals. Here you can see spikes from particulate material causing high variance and incorrect retreival of the SIA shape. #. Apply the dialog. Clicking *Apply* will disable the SIA shape option in the **Limit Options Dock** and instead use the per-mass SIA retrieved previously.