Non-targeted Analysis Using an ICP-ToF ====================================== The simultaneous acquisition of the full elemental mass range makes ICP-ToF an excellent particle screening tool. In this example, elements of interest in a soil extract are automatically selected using parameters set during import. A full description of the non-target screening tool is available at :ref:`Non-target Screening`. #. Download the required data file. The ``tof_non_target`` Nu directory can be found in the `example_3_data.zip `_ archive. #. Import the sample file. Open the sample file to start the :ref:`Data Import` wizard. Select *Non-target screening* and enter 1000 into the *Screening ppm* field, this will select elements with greater than 1000 particles per million events. Press *Ok* to start the screening process. #. Verify element selection. .. _nontarget_dialog: .. figure:: ../images/tutorial_tof_nontarget.png :width: 60% :align: center Elements selected during the screening are colour coded from black (low) to red (high) to show the relative number of detected particles. The non-target screening will automatically select all elements with detection counts above the given *Screening ppm* value, in this case Fe, La and Ce. Press *Ok* to import the data.